Material selection and design analysis of multi-purpose disposable safety syringe
-
2018-10-09 https://doi.org/10.14419/ijet.v7i4.13.22985 -
Abstract
Current design of safety syringes requires two handed operation and additional processes which is not similar to the normal syringes. Due to this concern, a new design of safety syringe is introduced in order to produce a safety syringe which allows a single-handed operation and similar to the operation of a normal syringes. This paper presents the material selection process and design analysis of a newly developed multi-purpose disposable safety syringe. Based on the design analysis, the force which needed to dismantle the nozzle is found to be 20 N and this value is practical for the end users. The finite element analysis had also shown that the design concept is safe and has safety feature for the user to use. In addition, copolymer is proven as the best material selection for safety syringe production.
-
References
[1] Levin MV (2006), A new generation of disposable syringes. Biomedical Engineering 40(3), 140–143
[2] Sohn JW, Kim BG, Kim SH & Han C (2006), Mental health of healthcare workers who experience needlestick and sharps injuries. Journal of Occupational Health 48(6), 474–479
[3] Alert N (1999), Preventing needlestick injuries in health care settings. DHHS (NIOSH) Publication
[4] Canadian Centre for Occupational Health and Safety (2014), Needlestick and sharps injuries: OSH Answers. http://www.ccohs. ca/oshanswers/diseases/needlestick_injuries.html
[5] Steinglass R, Boyd D, Grabowsky M, Laghari AG, Khan MA, Qavi A & Evans P (1995), Safety, effectiveness and ease of use of a non-reusable syringe in a developing country immunization programme. Bulletin of the World Health Organization 73(1), 57–63
[6] Thye SL & Bakri R (2007), Retractable needle syringe. Ministry of Health Malaysia.
[7] Hsieh H, Ho C & Wang S (2005), Safety Syringe. Syriteck Medical Devices Co., Ltd. Patent No. US 6979314 B2
[8] Van Dyke LR (2002), Automatically Retractable Needle Safety Syringe. Patent No. US6413236 B1
[9] Villanueva G (1999), Safety Syringe. Patent No. US 5938641 A
[10] Liu WN (1999), Safety Syringe with Retractable Standard Needle. Patent No. US 5899887 A
[11] Lee R (1994), Simple Retractable Safety Syringe. Patent No. US 5344403 A
[12] Walsh A (2010), Retractable Safety Syringe with Compressible Gasket Means. Global Medisafe Holdings Limited. Patent WO 2010096879 A1
[13] Mohd Zabidi SZ (2010), Design and Development of Safety Syringe. Thesis, Universiti Putra Malaysia.
[14] Basri AA (2011), Safety Syringe Product Development. Thesis, Universiti Putra Malaysia.
[15] Abu Talib AR, Basri AA, Mohd Zabidi SZ, Yahaya MS & Mat Isa K (2011), Safety Syringe. Selia-Tek Medical Sdn.Bhd, Universiti Putra Malaysia. Patent No. US 20140171866 A1.
[16] Ishak MN, Abu Talib AR, Harmin MY & Zahari R (2016), Design improvement of multi-purpose disposable safety syringe. World Research & Innovation Convention on Engineering & Technology
[17] Phase CD & Step CE (2013), Conceptual Design Phase: Concept Evaluation Step, 1–40
[18] Oberdorfer PE (2000), Syringes - Latex and Silicone Oil in the Design and Manufacture of Disposable Syringes, 1–3
[19] Brendel CM (2009), Biocompatibility of polymer implants for medical applications. http://etd.ohiolink.edu/send-pdf.cgi/Brendel Christopher M.pdf?acc_num=akron1246892895
[20] Emco Industrial Plastics Inc. (2016), PP - Homopolymer. http:// www.emcoplastics.com/pp-homopolymer/
[21] Sepe M (2012), The Online Magazine of Plastics Technology: Plastics Technology. How Do You Like Your Acetal: Homopolymer or Copolymer? : Plastics Technology. http://www.ptonline.com/ columns/how-do-you-like-your-acetal-homopolymer-or-copolyme
[22] Richard Claudio AH (2004), Injection Pressures by Anesthesiologists During Simulated Peripheral Nerve Block. New York: US National Library of Medicine National Institutes of Health.
-
Downloads
-
How to Cite
Ishak, M. N., Talib, A. R. A., & Harmin, M. Y. (2018). Material selection and design analysis of multi-purpose disposable safety syringe. International Journal of Engineering & Technology, 7(4.13), 214-220. https://doi.org/10.14419/ijet.v7i4.13.22985Received date: 2018-12-02
Accepted date: 2018-12-02
Published date: 2018-10-09